PM-240-MTP Multifiber Optical Power Meter
The PM-240-MTP optical power meter is designed to measure absolute or relative optical power in optical networks terminated with 12/241 multifiber MTP/MPO connectors, in both SM and MM fibers. The tester can measure simultanously optical power level in all up to 12 fibers of MTP/MPO connectors, it can recognize “live” and “dark” fibers. It eliminates the need of fan-out from multi to single fiber connectors. Together with LS-240-MTP light source can measure Insertion loss in all 12 fibers at same time, in addition it is able to to check the polarity status of fibers interconnection between both MTP connectors. The input port ensures interconnection between measured MTP/MPO connector and 12/24 photodetectors. Due to physical contact between connectors the cleanness of both is required, preferably the visual check before measurement is suitable.
The internal memory allows measurement storage and uploading of more than 500x 12-fibers cables including cable and fiber number, wavelength, absolute value or insertion loss. The Data Exporter software allows the user to export stored data to Excel sheet, or other applications.
The rechargeable Li-Pol battery ensures long term operation with a minimum service life of 2 years. Batteries can be charged via a USB port.
- Hand held, light weight
- Easy measurement of multifiber MTP/MPO connectors, no Fanout required
- Measures power in both SM and MM fibers
- Optical Power and Optical Loss measurement
- Large display with backlight
- Table view or Bar view of all 12/241 fibers
- Polarity test (with LS-240-MTP Light Source)
Parameter: |
|
Note: |
Photodetector |
InGaAs |
|
Working wavelengths |
850, 1300, 1310, 1490, 1550, 1625 nm |
can be customized |
Dynamic range: |
-50 dBm to +10 dBm -45 dBm to +10 dBm |
1300, 1310, 1490, 1550, 1625 nm 850 nm |
|
|
|
Photodetector |
Si |
|
Working wavelengths |
850 nm |
|
Dynamic range |
-40 dBm to +10 dBm |
850 nm |
|
|
|
Scan time |
3 sec |
12 fibers |
Uncertainty |
± 15% |
1310, 1550 nm @ -20 dBm |
Resolution |
0.01 |
|
AWD/Modulation Detection |
-40 dBm -35 dBm |
1300 – 1625 nm 850 nm |
Dimensions |
185 x 100 x 45 mm |
W x H x D |
Weight |
0.5 kg |
|
Temperature operating storage Humidity (non condensing) |
-10 to +50 °C -40 to +70 °C 0 – 95% |
|
Battery working time |
> 300 hrs |
|
Battery life time |
> 2 years |
3000 mAh Li-Pol |
Compliant with RoHS-requirements (2002/95/EG, 27.01.2003)
Options:
• Master MTP/MPO connector patchcord
PM-240 |
- |
MTPxx |
|
|
M |
- |
xx |
|
|
|
|
|
|
|
|
|
|
|
|
fibers |
|
– male, pins2 |
|
- |
InGaAs photodetector |
|
|
|
xx |
12/24 |
|
|
|
SI |
Si photodetector |
Note: 2) standard: input MTP connector with pins – male